Portable Conductivity Meter
+
  • Portable Conductivity Meter

Portable Conductivity Meter

DD-30

The Portable Conductivity Meter is a compact and portable laboratory-grade instrument designed for precise conductivity measurements. It is widely applicable in industries such as petrochemicals, biopharmaceuticals, wastewater treatment, and mining/metallurgy, as well as for measuring conductivity, resistivity, salinity, and TDS (Total Dissolved Solids) in aqueous solutions.

  • Description
  • Accessories
    • Brand: WATENV
    • Commodity name: Portable Conductivity Meter
    • 系列: DD-30

    Features

    Stores up to 500 datasets with options to delete, review, and export.

    Advanced Functions

    Power-off protection, factory reset, auto-shutdown, and parameter memory.

    Automatically recognizes four standard conductivity solutions: 146.6 μS/cm, 1408 μS/cm, 12.85 mS/cm, and 111.3 mS/cm.

    IP65-rated enclosure protects against liquid splashes and corrosion.

    Supports continuous measurement mode and stable measurement mode (DD30-3/4);

    Supports single-point and multi-point electrode calibration function (DD30-3/4);

    Supports manual temperature compensation and automatic temperature compensation (DD30-3/4).

     

    Technical Parameters

    ModelDD30-1DD30-2DD30-3DD30-4
    Accuracy Class1.01.01.00.5
    Power SupplyInput AC100~240V 50~60Hz 0.35A; Output DC5V/2A
    Screen4-inches LCD
    Temperature CompensationManualManual

    Manual

    /Automatic

    Manual

    /Automatic

    Manual Temperature Compensation

    Range

    0.0~60.0℃0.0~60.0℃0.0~60.0℃0.0~60.0℃

    Automatic Temperature Compensation

    Range

    //-5.0~110.0℃-10.0~135.0℃
    Minimum Temperature Resolution0.1℃/℉0.1℃/℉0.1℃/℉0.1℃/℉
    Temperature Accuracy//

    ±0.4℃

    (0.0℃~60.0℃)

    ConductivityMeasurement Range0.00μS/cm~100mS/cm0.00μS/cm~200mS/cm0.000μS/cm~2000mS/cm
    Minimum Resolution0.01μS/cm0.01μS/cm0.01μS/cm0.01μS/cm
    Electronic Unit Error±1.0%(FS)±1.0%(FS)±1.0%(FS)±0.5%(FS)
    Electronic Unit Repeatability0.33%(FS)0.33%(FS)0.33%(FS)0.17%(FS)
    Instrument Error±1.5%(FS)±1.5%(FS)±1.5%(FS)±0.8%(FS)
    Instrument Repeatability0.7%(FS)0.7%(FS)0.7%(FS)0.4%(FS)
    TDSMeasurement Range/0.00mg/L~100g/L0.000mg/L~1000g/L
    Minimum Resolution/0.01mg/L0.01mg/L0.001mg/L
    Electronic Unit Error/±1.0%(FS)±1.0%(FS)±0.5%(FS)
    ResistivityMeasurement Range///5.00-100.0MΩ.cm
    Minimum Resolution///0.01Ω.cm
    Electronic Unit Error///±0.5%
    salinityMeasurement Range///0.00-8.00%
    Minimum Resolution///0.01%
    Electronic Unit Error///±0.1%
    InstrumentError///±0.2%
    Calibration Points111-31-3
    Standard Solution Recognition

    146.6μS/cm, 1408μS/cm

    12.85mS/cm

    146.6μS/cm, 1408μS/cm

    12.85mS/cm, 111.3mS/cm

    Data Storage200 sets200 sets500 sets1000 sets
    Standard ElectrodeLC-DJS-1(Platinum Black)

    LC-DJS-1(Platinum Black)

    T-10K

    Electrode InterfaceQ9Q9Q9Q9
    Dimensions (mm)200×80×50200×80×50200×80×50200×80×50
    Weight (kg)0.450.450.450.45
  • pH Electrode

    Combination

    pH Electrode

    Glass

    ORP ElectrodeConductivity ElectrodeTemperature Electrode

     

    Conductivity Standard Solution

    Nominal Value

    Specification

    Shelf Life

    146.6 µs/cm

    4 × 50 mL bottles

    6 months

    1408 µs/cm

    4 × 50 mL bottles

    12 months

    12.85 ms/cm

    4 × 50 mL bottles

    12 months

    111.3 mS/cm

    4 × 50 mL bottles

    12 months

     

    pH Electrode Combination

    E-201-C

    E-301-C

    pH Measurement range

    0~14ph

    0~14ph

    Junction Type

    Single Salt Bridge

    Single Salt Bridge

    Reference System

    Ag/AgCl

    Ag/AgCl

    Reference Junction Type

    Non-refillable

    Non-refillable

    Connector Type

    BNC (Q9)

    BNC (Q9)

    Cable Length

    1 meter

    1 meter

    Operating Temperature

    5~60℃

    5~60℃

    Electrode Storage Solution

    3 mol/L KCl

    3 mol/L KCl

    Sensing Membrane

    Spherical glass bulb

    Spherical glass bulb

    Combination

    Reference + sensing

    Reference + sensing + temperature probe

    Temperature Electrode

    /

    2.5mm audio jack connector

    Dimensions

    12mm*160mm

    12mm*160mm

    Weight

    0.060kg

    0.078kg

     

    pH Electrode Glass

    6502

    6503

    pH Measurement Range

    2–12

    0–14

    Junction Type

    Ceramic

    Ceramic

    Electrode Body Material

    Glass

    Glass

    Reference System

    Ag/AgCl

    Ag/AgCl

    Reference Junction Type

    Double junction, Refillable

    Double junction, Refillable

    Connector Type

    Q9

    Q9

    Cable Length

    1 meter

    1 meter

    Operating Temperature

    5–90°C

    5–90°C

    Electrode Storage Solution

    3 mol/L KCl

    3 mol/L KCl

    Sensing Membrane

    Spherical glass bulb

    Spherical glass bulb

    Combination

    Reference + sensing

    Reference + sensing

    Dimensions

    φ12 × 165 mm

    φ12 × 165 mm

    Weight

    0.1 kg

    0.1 kg

    Electrode Selection Guide

    pH Electrode Glass

    6502

    6503

    General Laboratory Use

    Y

    Y

    Strong Acids/Bases or High-Corrosion Environments

    ×

    Y

    Low-Conductivity Solutions

    Y

    ×

    Pure Water

    Y

    ×

    Biochemical samples

    ×

    ×

    Tris-containing solutions

    ×

    ×

    Soil Water, Wastewater

    ×

    Y

    Viscous, Emulsion, or Oily Substances (emulsions, food products, cosmetics, pigments, paints)

    ×

    Y

    (Note: Y = Suitable, × = Unsuitable)

     

    ORP Electrode

    5001

    Body Material

    Plastic

    Sensor Design

    Needle-Type PlatinumTip

    Reference System

    Plastic Fiber, Single Junction, Non-Refillable

    Connector Type

    Q9

    Cable Length

    1 Meter

    Operating Temperature

    5–60°C

    Electrode Storage Solution

    3mol/L KCl Solution

    Electrode Features

    General-Purpose

    Dimensions

    Φ12×160mm

    Weight

    0.1kg

     

    Conductivity Electrode

    DJS-0.1

    DJS-1 (Platinum Black)

    DJS-10 (Platinum Black)

    Application

    Ultra-pure water, pure water

    Environmental water, routine water

    High-electrolyte solutions

    Housing Material

    Glass

    Glass

    Glass

    Cell Constant (K)

    0.1±0.02

    1±0.2

    10±2

    Measurement Range

    0.055~300µS/cm

    50~2000µS/cm

    2~100mS/cm

    Dimensions (mm)

    φ12×155

    φ12×155

    φ12×155

    Weight (kg)

    0.1

    0.1

    0.1

     

    Temperature Electrode

    T-10K

    Material

    Stainless steel

    Temperature Coefficient

    10K

    Measurement Range

    0-100℃

    Dimensions

    φ12×155 mm

    Weight

    0.1kg